Nuclear Magnetic Resonance

Level 3 · 11 min read

Nuclear magnetic resonance is the absorption and re-emission of radio waves by nuclei in a magnetic field, at the frequency where the nuclei precess; the frequency reveals the field and the chemical environment.

On this page

  • Colour = topic mix
  • Classical mechanics
  • Discrete math
  • Calculus
  • Electromagnetism
  • Linear algebra
  • Probability
  • ReadReady to readNot yet

TL;DR

Nuclear magnetic resonance, NMR, is the way nuclei with spin respond to radio waves when they sit in a magnetic field. The nuclei precess at a frequency set by the field, and a radio pulse at that frequency tips them enough to produce a signal that a coil can detect. Because the exact frequency depends slightly on the chemical environment of each nucleus, the signal tells chemists which atoms are present and how they are arranged, and it also underlies magnetic resonance imaging.

Builds on: Nuclear Spin, Larmor Precession, Resonance, Boltzmann Distribution · Leads to: Bloch Equations

Where it's used

Identifying molecules in chemistry. Chemists dissolve a sample, place it in a strong magnet and record its NMR spectrum. The spectrum shows peaks at different frequencies, and the pattern of peaks reveals which hydrogen and carbon atoms are present and which neighbours they have. It is a routine tool for confirming that a newly made compound is the one intended.

Determining protein structures. Biologists use NMR to measure how atoms in a protein sit relative to one another in solution, without first growing a crystal. Kurt Wüthrich shared the 2002 Nobel Prize in Chemistry for his NMR studies of biological macromolecules in solution.

Magnetic resonance imaging. A scanner uses the same physics on the water and fat in a body, adding magnetic-field gradients so that the frequency of each signal indicates where it came from. The 2003 Nobel Prize in Physiology or Medicine recognised Paul Lauterbur and Peter Mansfield for developing this imaging method. This lesson explains the physics only; it is not guidance about any medical test.

How it works

Follow one hypothetical sample of spin-half nuclei with a Larmor constant γ = 2 rad per second per tesla, placed in a field B = 3 tesla. The numbers are chosen to keep the arithmetic short.

  1. The nuclei precess. Each nucleus's magnetic moment circles the field at ω = γB = 2 × 3 = 6 radians per second. Check: the rate does not depend on how the nucleus is tilted.
  2. The sample leans slightly. Most nuclei have their spin along the field, but a tiny excess points along it, and that excess makes the sample's net magnetization point along the field. Check: without the field there would be no net direction to detect.
  3. Tip the magnetization with a pulse. A short radio pulse at the same frequency, 6 radians per second, is applied sideways. Because the pulse is at the precession frequency, its effect keeps building up as it acts. Check: a pulse at 5 radians per second would keep changing direction relative to the precession and would tip the magnetization much less.
  4. Rotate it into the sideways plane. A pulse that tips the magnetization by 90 degrees leaves it lying sideways, where it can rotate. Check: that sideways magnetization now precesses at 6 radians per second.
  5. Detect the rotation. A coil next to the sample sees the rotating magnetization as a voltage that oscillates at 6 radians per second. This is the same principle as Faraday's law: a changing magnetic flux through a loop induces a voltage. Check: after 0.26 seconds, about π/12, the phase has moved by 6 × 0.26 ≈ 1.57, which is a quarter turn.
  6. Turn the signal into a spectrum. The recorded voltage is a sum of oscillations, one per nucleus. A Fourier transform separates them into peaks at their frequencies. Check: a single nucleus gives one peak at 6 radians per second.

The plot shows the same resonance from the other side: the response is largest when the pulse frequency matches the precession frequency and falls off on either side.

02468101200.20.40.60.81pulse frequency (rad per second)response
  • response to a pulse at frequency x
The response peaks where the pulse frequency equals the precession frequency, 6 radians per second in this example. Its width depends on how quickly the nuclei lose their alignment.

The math

Definition

Definition: Nuclear magnetic resonance

A nucleus with spin II and gyromagnetic ratio γ\gamma in a magnetic field of strength BB has energy levels Em=−γℏB mE_m = -\gamma\hbar B\, m, where mm takes the values −I,−I+1,…,I-I, -I+1, \ldots, I. The radio photon that drives a transition between neighbouring levels has angular frequency ω=γB\omega = \gamma B. That is the NMR resonance condition.

Step 1 in symbols: the levels

A spin-half nucleus has I=1/2I = 1/2, so m=±1/2m = \pm 1/2 and there are two levels. Their separation is

ΔE=E−1/2−E+1/2=γℏB.\Delta E = E_{-1/2} - E_{+1/2} = \gamma\hbar B.

For the sample in the walkthrough, γB=6\gamma B = 6 rad per second, so ΔE=ℏ⋅6\Delta E = \hbar \cdot 6.

Step 2 in symbols: the population

The ratio of the two populations is

NlowNhigh=exp⁡ ⁣(ΔEkBT),\frac{N_{\text{low}}}{N_{\text{high}}} = \exp\!\left(\frac{\Delta E}{k_B T}\right),

where kBk_B is the Boltzmann constant and TT is the temperature. When ΔE\Delta E is much smaller than kBTk_B T, the excess in the lower level is tiny. This is why the net magnetization in Step 2 of the walkthrough is so small.

Step 3 in symbols: the pulse and the response

A pulse at frequency ω\omega tips the magnetization by an angle that grows most when ω=ωL\omega = \omega_L. A simple model of the response is the peaked form in the plot,

R(ω)=11+(ω−ωLΓ)2,R(\omega) = \frac{1}{1 + \left(\dfrac{\omega - \omega_L}{\Gamma}\right)^2},

where Γ\Gamma sets the width of the peak. The width is set by how quickly the nuclei lose their alignment, which is the subject of the Bloch Equations.

Step 5 in symbols: the signal

Let M\mathbf{M} be the net magnetization of the sample. Its component in the sideways plane rotates at the precession frequency, M⊥(t)∝e−iωLtM_\perp(t) \propto e^{-i\omega_L t}, with ωL=γB\omega_L = \gamma B. A coil picks up the changing flux, so the induced voltage is proportional to ωLM⊥\omega_L M_\perp. A stronger field therefore gives a larger signal from the same sample, and it also raises the frequency.

Worked examples

Example
Example 1: the period of precession (tiny)

A nucleus with γ = 2 rad per second per tesla sits in a field of B = 2 tesla.

  1. Compute the precession frequency: ωL=γB=2×2=4\omega_L = \gamma B = 2 \times 2 = 4 rad per second.
  2. Compute the period, the time for one full turn: T=2π/ωL=2π/4=π/2T = 2\pi / \omega_L = 2\pi / 4 = \pi/2.
  3. Evaluate: π/2 is about 1.57 seconds. Check: 4 times 1.57 is about 6.28, which is 2π.
  4. Interpretation: a pulse must be applied at 4 radians per second to tip this nucleus, and the signal repeats every 1.57 seconds.
Example
Example 2: how big is the signal? (standard)

A proton in a 1.5 tesla scanner has ω = 4.01 × 10^8 rad per second, the value from the Larmor lesson, which corresponds to a frequency of about 63.8 MHz. Compare the energy gap with the thermal energy at room temperature, T = 300 K.

  1. Compute the photon energy at the resonance frequency: E=hνE = h\nu, with h=6.626×10−34h = 6.626 \times 10^{-34} J s and ν=6.38×107\nu = 6.38 \times 10^7 Hz. The product is 4.23×10−264.23 \times 10^{-26} J.
  2. Compute the thermal energy: kBT=1.381×10−23×300=4.14×10−21k_B T = 1.381 \times 10^{-23} \times 300 = 4.14 \times 10^{-21} J.
  3. Divide: ΔE/(kBT)=4.23×10−26/4.14×10−21≈1.0×10−5\Delta E / (k_B T) = 4.23 \times 10^{-26} / 4.14 \times 10^{-21} \approx 1.0 \times 10^{-5}.
  4. Interpret with Step 2: the population ratio is exp⁡(1.0×10−5)≈1+1.0×10−5\exp(1.0 \times 10^{-5}) \approx 1 + 1.0 \times 10^{-5}. The lower level holds only about one part in 100,000 more nuclei than the upper level.
  5. Consequence: almost every nucleus contributes nothing to the net signal, and the spectrometer must average many scans to see a usable signal. This is why NMR needs a large sample and a strong magnet.
Example
Example 3: a chemical shift in a spectrum (realistic)

A chemist runs a proton NMR spectrum on a 400 MHz spectrometer, where the reference peak sits at exactly 400 MHz. A peak appears 1200 Hz above the reference. The position of a peak relative to the reference, in parts per million, is called its chemical shift.

  1. Write the definition: chemical shift in ppm is the difference from the reference frequency, divided by the reference frequency, times one million.
  2. Substitute the difference: 1200 Hz.
  3. Divide by the reference frequency: 1200/(400×106)=3.0×10−61200 / (400 \times 10^6) = 3.0 \times 10^{-6}.
  4. Multiply by one million: 3.0 ppm.
  5. Interpretation: the nucleus is sitting in a slightly different local field because of the electrons around it. The shift of 3.0 ppm is a fingerprint of that environment, which the chemist compares with tables of known values.
Example
Example 4: the same peak on a different spectrometer (try it)

A second spectrometer runs at 600 MHz. The same nucleus in the same molecule appears 1800 Hz above its reference. Compute the chemical shift in ppm.

  1. Divide the frequency difference by the reference frequency: 1800 divided by 600 million.
  2. Multiply the result by one million.

Try it: work both steps before reading on.

Answer: 1800 / (600 × 10^6) = 3.0 × 10^-6, which is 3.0 ppm. The shift in ppm is the same as on the 400 MHz instrument, even though the raw frequency offset is different. This is why chemists report shifts in ppm rather than hertz.

Common mistakes and tradeoffs

"The strong field lines all the nuclei up, so the sample is fully polarised." It is tempting to think a magnet aligns everything. In fact the excess is about one part in 100,000 at room temperature, as the calculation above shows. The signal comes from that small excess and from the coil's sensitivity to changing flux.

"Resonance means the sample absorbs any radio frequency." The response is large only near the precession frequency. The peak has a finite width, set by how quickly the nuclei lose their alignment, so frequencies close to the peak still respond a little.

"A bigger magnet is always better." A stronger field increases the population excess and the signal, but it also raises the cost and the engineering needed, and it changes the frequency the electronics must handle. The gain is real but has to be weighed against those costs.

"NMR and MRI are the same measurement." They share the physics. An MRI scanner adds magnetic-field gradients so that the frequency of each signal indicates where in the body it came from. A chemist's spectrum has no gradients and reports chemical environments instead.

When to use NMR: when you need to identify a molecule's structure or the arrangement of atoms in solution, and you can supply a reasonable amount of sample. When not to use it: when the sample is tiny, because mass spectrometry needs far less material, or when a crystal can be grown, because X-ray crystallography gives atomic positions more directly.

MethodStrengthLimitation
NMRAtomic environments, structure in solutionLow sensitivity, needs a large sample
Mass spectrometryVery small samples, molecular massesLess direct about arrangement
X-ray crystallographyPrecise atomic positionsNeeds a good crystal

History

Isidor Rabi developed the resonance method for nuclear moments with molecular beams in the late 1930s, and he received the 1944 Nobel Prize in Physics for it. In 1946 Felix Bloch and Edward Purcell, working independently, observed NMR in bulk matter, which made the method usable for solids and liquids. Both received the 1952 Nobel Prize in Physics. Chemists began using the chemical shift to identify molecules in the 1950s. Richard Ernst's introduction of Fourier-transform NMR, which recorded all frequencies at once, made spectra far faster to collect, and earned him the 1991 Nobel Prize in Chemistry. Imaging followed in the 1970s, when Lauterbur and Mansfield developed the field-gradient methods that made MRI possible.

Recap

  • Nuclei with spin precess in a magnetic field at the Larmor frequency, ω = γB.
  • A tiny excess of nuclei along the field, given by the Boltzmann ratio, makes a net magnetization.
  • A pulse at the precession frequency tips the magnetization into the sideways plane, where it rotates and induces a voltage.
  • A Fourier transform of that voltage gives a spectrum, and chemical shifts in ppm identify chemical environments.
  • Read next: Bloch Equations for how the signal decays, and Faraday's Law of Induction for how the coil detects it.

Check yourself

QuestionState the NMR resonance condition for a nucleus with gyromagnetic ratio γ in a field B.Show answer

The radio frequency must equal the precession frequency, ω = γB, which is also the energy gap ħω between the two spin levels divided by ħ. A pulse at this frequency drives the transition.

QuestionA nucleus has γ = 3 rad per second per tesla in a field of B = 4 tesla. What is its angular precession frequency, and what is the period of one turn?Show answer

ω = 3 × 4 = 12 rad per second. The period is T = 2π / 12 = π/6, which is about 0.52 seconds. Check: 12 times 0.52 is about 6.3, which is 2π.

QuestionWhy does an NMR experiment need a large sample and many repeated scans, even with a strong magnet?Show answer

The population excess that makes the net magnetization is about one part in 100,000 at room temperature, so most nuclei contribute nothing net. The signal is small. Averaging many scans helps because the signal adds up in step with the number of scans, while random noise grows only as its square root, so the signal-to-noise ratio improves as the square root of the number of scans.

QuestionA spectrum on a 300 MHz instrument shows a peak 900 Hz above the reference. What is its chemical shift in ppm?Show answer

900 / (300 × 10^6) = 3.0 × 10^-6, which is 3.0 ppm. Check: the same shift on a 600 MHz instrument would appear at 1800 Hz above the reference, because the frequency offset scales with the field.